Senba Sensing

Senba SensingStock Code:300701Focus on PIR R&D manufacturing over 19years

Senba Sensing

+86-755-82594732
Pressure Transmitter

Pressure Transmitter

Home > Products > Pressure Transmitter

Model 178

Pressure Transmitter

Differential pressure sensor/transmitter

Pressure medium: air or other non-conductive gas

Accuracy: ±1.0%,±0.4%,±0.25%

Range: unidirectional: 25 Pa to 25000 Pa/bidirectional: ±10 Pa to ±10000 Pa

Signal output: 4-20mA, 0-5Vdc, 0-10Vdc

Overload: Forward and reverse 15PSL (100KPa)

Model 178
General Description

The core component of the Model 178 series differential pressure
sensor/transmitter is a capacitive pressure sensor, which consists of a
stainless steel diaphragm with a fixed electrode forming a capacitor
whose value varies with pressure. Using the company's special patented
technology carefully designed sensor sensors and unique detection
circuit to ensure good temperature performance and long-term stability
of the product.

The Model 178 differential pressure
sensor/transmitter can be used to measure differential or gauge pressure
with outputs of 0-5V, 0-10V, 4-20mA, and can also be customized for
other outputs. The accuracy of Model 178 was ±1.0%FS, 0.4%FS or 0.25%FS
at room temperature. The temperature compensation ranges from 4 ℃ to
77℃, and the temperature effect is less than ±0.05%FS/℃.

Model
178 has excellent performance-price ratio, widely used in HVAC,
environmental pollution control, clean engineering, medical instruments
and equipment, oven pressurization and furnace pressure control, natural
gas, gas pipe network monitoring, underground ventilation and power
plant wind pressure monitoring and other fields.

Features

Strong aluminum case, according to NEMA 4/IP65 standard

3-1/2 bit LCD display

SS304 Embedded installation

The positive/negative pressure end can withstand at least 100kPa overload without damage

One key zero button

Full protection of misconnection

Environmental protection design, in line with ROHS standards

Minimum pressure range of 10Pa!

Technical Data
ParameterStandardSpecial selectionSpecial selection
Accuracy RSS*(under constant temperature)±1.0%FS±0.40%FS±0.25%FS

nonlinearity(best fitting line)

±0.97%FS±0.37%FS±0.24%FS
lag±0.20%FS±0.10%FS±0.05%FS
non-repeatability±0.10%FS±0.10%FS±0.05%FS
Temperature effect
Zero point/Full range offset(/℃)±0.05%FS±0.03%FS±0.02%FS
Compensation range4℃ to 77℃
Maximum static pressure15PSI(100kPa) 
OverloadBoth positive and negative are 15PSI(100kPa)
Stability ratio±0.5% FS/YR
Preheat5 seconds to reach the performance specification
Installation position effectWhen factory calibration, the sensor is placed vertically. In order to ensure accuracy, necessary fine tuning should be made when installing other positions at zero point


Environmental and mechanical parameters
Operating temperature0℃ to 60℃
Storage temperature-40℃ to 85℃
Electrical connectionPluggable terminal with PG-9 cable locking device
Pressure connection3/16"、φ8、φ6.5 brass tower pressure connection(optional)
Output regulation
ress the zero button to clear zero
Pressure mediumAir or other non-conductive gas
ShellDie cast aluminum, NEMA 4/IP65
Weight330g
InstallationThere are two 0.165 "diameter mounting holes in the housing. M4 X 12 pan head screws with washers and nuts are included in both cases, as shown in figure for appearance and dimensions. Stainless steel mounting plates with lugs are optional